Chapter 2 Performance Tests

58502A Calibration (Option 010 Only)

User’s Guide

2-19

2

Use the terminal and serial port to send one of the following SCPI
commands to incrementally adjust the EFC control signal to bring the
oscillator frequency closer to 10 MHz:

DIAG:CAL:ROSC:EFC:ABS UP (or)
DIAG:CAL:ROSC:EFC:ABS DOWN

Where UP increases frequency and DOWN decreases frequency.

3

As you approach the proper frequency, change the step size to smaller
values by sending the following SCPI command:

DIAG:CAL:ROSC:EFC:ABS:STEP <value>

The default step size is 10,000. A recommended first reduction in size
should be to 1,000.

4

Continue to adjust UP or DOWN with finer increments until the
frequency output is 10 MHz

±

<

1

×

10

9

(0.01 Hz).

5

Record the date, EFC value, and frequency value on the Calibration
recorded, located on page 2-20.

6

Return the EFC step size to its original value by sending the following
SCPI command:

DIAG:CAL:ROSC:EFC:ABS:STEP 10000

This completes the 58502A calibration.