Konica Minolta Chroma Meter CS-200 User Manual

Page 110

Advertising
background image

108

De

scr

ipti

on

Operation Procedure

(With data management software CS-S10w)

Setting Necessary for Object Color Measurement

1.

Set one or more tungsten lumps
or equivalent as illumination

source toward white calibration

plate as in the right illustration.

• Set this instrument vertical
to white calibration plate.
• Keep the angle between
illumination light source and
white calibration plate at 45˚.

2.

Start up this instrument, PC, and software CS-S10w.

White Calibration

3.

Set color measuring mode for object color with CS-S10w.

4.

Conduct white calibration with CS-S10w.

Measurement of Object

5.

Set object on the same position and at the
same angle as white calibration plate.

6.

Now, measurement can be done with CS-S10w.

See instruction manual of data management software CS-S10w for details.

Illumination

Light Source

Illumination

Light Source

White calibration plate

Advertising