Specification, De scr ipti on – Konica Minolta Chroma Meter CS-200 User Manual

Page 122

Advertising
background image

120

De

scr

ipti

on

Specification

Item

CHROMA METER CS-200

Measurement range

0.01 - 200,000 cd/m

2

(Measuring angle 1˚)

0.01 - 5,000,000 cd/m

2

(Measuring angle 0.2˚)

0.01 - 20,000,000 cd/m

2

(Measuring angle 0.1˚)

Accuracy
(Measuring angle 1˚)∗1
(Temperature: 23˚±2˚,
RH 65% max.)

150 cd/m

2

(for illuminant A) L

v

±2 %±1digit

xy ±0.002

0.01 - 0.5 cd/m

2

(for illuminant A) L

v

±0.02 cd/m

2

±1digit ---

0.5 - 1 cd/m

2

(for illuminant A) L

v

±0.02 cd/m

2

±1digit xy ±0.007

1 - 10 cd/m

2

(for illuminant A) L

v

±2 %±1digit

xy ±0.004

10 - 200,000 cd/m

2

(for illuminant A) L

v

±2 %±1digit

xy ±0.003

(5,000 cd/m

2

for illuminant A) color filter (R,G,B)

xy ±0.006

Repeatability
(Measuring angle 1˚)∗2

0.01 - 1 cd/m

2

(for illuminant A) L

v

0.01 cd/m

2

+1digit --- (2 σ) (AUTO)

1 - 2 cd/m

2

(for illuminant A) L

v

0.5 %+1digit xy 0.002 (2 σ) (AUTO)

2 - 4 cd/m

2

(for illuminant A) L

v

0.5 %+1digit xy 0.001 (2 σ) (AUTO)

4 - 8 cd/m

2

(for illuminant A) L

v

0.5 %+1digit xy 0.0005 (2 σ) (AUTO)

8 - 200,000 cd/m

2

(for illuminant A) L

v

0.1 %+1digit xy 0.0004 (2 σ) (AUTO)

Measurement time

Automatically set between approx. 1 sec/meas. and 60 sec/meas. (AUTO)
Automatically set to approx. 1 sec/meas. or 3 sec/meas. (LTD.AUTO)
approx. 0.5 sec/meas.(Super-FAST) approx. 1 sec/meas.(FAST)
approx. 3 sec/meas. (SLOW)

approx. 12 sec/meas. (Super-SLOW)

Measurement method

Spectral method, Grating + linear photo diode array

Measuring angle

1˚, 0.2˚, 0.1˚ (Switchable)

Minimum measuring area

ø0.5 mm, ø0.1 mm (With close-up lens No.107)

Minimum measuring
distance

296 mm (Distance from front edge of metal lens barrel)

Color space

L

v

x y, L

v

u’ v’, L

v

T∆uv, XYZ, dominant wavelength

Measurement synchroniza-

tion frequency

Vertical synchronization frequency:40.00 to 200.00 Hz

Interface

USB1.1

Power source

AC adapter and 4 AA-Size Batteries

Battery life

3 hours approx. (in Continuous measurement/FAST mode, with AA-Size bat-

tery under company testing KONICA MINOLTA SENSING's conditions.)

Size

95 (W) x 127(H) x 334(L) mm

Weight

1.8 kg (without battery)

Operating temperature
/humidity range

0 to 40˚C, RH 85 % or less (at 35˚C) with no condensation

Storage temperature
/humidity range

0 to 40˚C, RH 85 % or less (at 35˚C) with no condensation

Standard accessory

Lens cap, Holding cap CS-A24, ND eyepiece filter CS-A27, AC adapter AC-

A23, Data management software CS-S10w Standard, USB cable (2 m) IF-A17

Optional accessory

Close-up lens No.107, Close-up lens No.122, ND filter (1/10) CS-A6,

ND filter (1/100) CS-A7, Calibration instruction (For ND filter),

Step up ring (40.5 to 55 mm) CS-A26, Angle finder V

N

, White calibration plate

(For 45-0) CS-A20, White calibration plate (For d-0) CS-A21, White calibra-

tion plate CS-A22, Soft case CS-A23, Data management software CS-S10w

Professional

∗1. 23˚C±2˚C, L

v

=0.01 to 10 cd/m

2

SLOW average of 30 measurements

L

v

= 10 cd/m

2

and higher SLOW average of 10 measurements

∗2. At 0.2˚ Measuring angle, the amount of received light is approx. 1/25 of that for 1˚.

Therefore, the repeatability becomes the same as that for 1˚ with 25 times lower luminance.

0.1˚ Measuring angle, the amount of received light is approx. 1/100 of that for 1˚.

Therefore, the repeatability becomes the same as that for 1˚ with 100 times lower luminance.

Advertising