Konica Minolta Chroma Meter CS-200 User Manual

Page 117

Advertising
background image

115

De

scr

ipti

on

Error

No.

Symptom

Item to Check

Corrective Action

Page to

Refer

9

Display of

the remaining

measurement

time freezes,

and measure-

ment is not

completed

for the set

measurement

time.

Is object to measure

stable?

When an object whose luminance
has changed greatly from the pre-
vious measurement is measured,
measurement may be interrupted to
set the optimum gain for measure-
ment and then remeasurement will
be performed; during remeasure-
ment, the remaining measurement
time display is frozen. Keep object to
measure stable.

Are you measuring an

object whose lumi-

nance has changed

greatly from the

previous measure-

ment?

When an object whose luminance
has changed greatly from the pre-
vious measurement is measured,
measurement may be interrupted to
set the optimum gain for measure-
ment and then remeasurement will
be performed; during remeasure-
ment, the remaining measurement
time display is frozen.

10

PC cannot

read output

data from this

instrument

when con-

nected with

USB. Com-

mand or data

cannot be

entered from

PC to this

instrument.

Has USB cable been

tightly connected?

Connect this instrument and PC
securely.

98

Hasn't remote mode

been cancelled?

Switch to remote mode by sending
connection command from PC to this
instrument.
Use standard accessory data manage-
ment software CS-S10w Standard.

99

Has prepared program

been correct?

Check referring to sample program. Use
standard accessory data management
software CS-S10w Standard.

-

11

Measurement
data or vari-
ous setting
becomes
invalid soon.

Hasn't backup battery
been insufficient after
a long time period of
nonuse? Has backup
battery been charged
enough at the time of
purchased?

Switch ON to charge backup battery.
Backup battery is to fulfill its func-
tion after fully charged for approx. 20
hours.
Shelf life of backup battery is 10
years under general use, but re-
placement is needed should memo-
ry disappears soon after full charge.
For replacing backup battery, please
contact the nearest KONICA MI-
NOLTA SENSING authorized service
facility.

4

12

Same error
message ap-
pears repeat-
edly.

Check corrective action
for error message.

If symptom does not improve, please
contact the nearest KONICA MI-
NOLTA SENSING authorized service
facility.

108