Enter, X ------ y, Lv 9 – Konica Minolta Chroma Meter CS-200 User Manual

Page 69: Lv 90.03, User cal ch id name ch copy ch reset

Advertising
background image

67

M

ea

su

re

m

en

t

Pr

ep

ara

tio

n

11.

Press

ENTER

key.

Calibration entry screen appears.

12.

Enter value for calibration.

key: 0 to 9 in ascending order. K, M,

decimal point and space available. If kept
pressed, value continuously changes.

key: 9 to 0 descending order. K, M,

decimal point and space available. If kept
pressed, value continuously changes.
K and M show x10

3

and x10

6

respectively.

(See p.71 for details on ‘Numerical value entry ranges’ )

13.

Press

key to move cursor

to second digit position.

14.

Repeat procedures from

12

to

13

as

necessary to enter calibration values

for Lv, x and y (or each value for

other color spaces) respectively.

Press

ESC key to stop.

15.

Press

ENTER

key.

After “ PLEASE WAIT...” appears,
<CH SETTING> screen appears.
When “ PLEASE WAIT...” appears, do not

U

S

E

R C

A

L

switch OFF. Doing so may break stored data.

16.

Press

ESC

key twice to return

to measurement screen.

<USER CAL DATA>
CH01:<DEFAULT>

Lv ------

cd/m

2

x ------

y ------

SET CAL VALUE.

OK : [ENTER]

[ SFT]  [ 2° ]

<

ANGL

>

1° VIEW

<

SPD

>

AUTO

<

SYNC

>

NO SYNC

<

LENS

>

STANDARD

<

MEM

>

AUTO NUM

<CURRENT DATA>

Lv 90.30

x 0.4210

y 0.4128

<

ANGL

>

1° VIEW

<

SPD

>

AUTO

<

SYNC

>

NO SYNC

<

LENS

>

STANDARD

<

MEM

>

AUTO NUM

<CURRENT DATA>

Lv 90.30

x 0.4210

y 0.4128

<USER CAL DATA>
CH01:<DEFAULT>

Lv 9-----

cd/m

2

x ------

y ------

SET CAL VALUE.

OK : [ENTER]

[ SFT]  [ 2° ]

<

ANGL

>

1° VIEW

<

SPD

>

AUTO

<

SYNC

>

NO SYNC

<

LENS

>

STANDARD

<

MEM

>

AUTO NUM

<CURRENT DATA>

Lv 90.30

x 0.4210

y 0.4128

<USER CAL DATA>
CH01:<DEFAULT>

Lv 9-----

cd/m

2

x ------

y ------

SET CAL VALUE.

OK : [ENTER]

[ SFT]  [ 2° ]

<

ANGL

>

1° VIEW

<

SPD

>

AUTO

<

SYNC

>

NO SYNC

<

LENS

>

STANDARD

<

MEM

>

AUTO NUM

<CURRENT DATA>

Lv 90.30

x 0.4210

y 0.4128

<USER CAL DATA>
CH01:<DEFAULT>

Lv 90.03

cd/m

2

x 0.4215

y 0.4130

SET CAL VALUE.

OK : [ENTER]

[ SFT]  [ 2° ]

<

CH SETTING

>

CH

01

:

<

DEFAULT

>

USER CAL

CH ID NAME

CH COPY

CH RESET

BREAK:[ESC]

[ SFT]  [ 2° ]

CH

01

:

<

DEFAULT

>

Lv ------

cd/m

2

x ------

y ------

<

MODE

>

SINGLE

[ SFT ]  [ 2° ]

<

ANGL

>

1° VIEW

<

SPD

>

AUTO

<

SYNC

>

NO SYNC

<

LENS

>

STANDARD

<

MEM

>

AUTO NUM

<

MEMORY DATA

>

M000 :(NO DATA)

Lv ------

x ------