63

COMP

ARING

Comparing the results:

When the lab tests your blood,
only the serum or plasma portion is
used. Your OneTouch

®

Profile

®

Meter uses only whole blood.

Comparing serum or plasma to
whole blood is like comparing a
temperature of 98.6˚F with the
temperature of 37˚C. To see if the
values are equal, you have to
adjust one of the results.

Whole blood results are approxi-
mately 12% lower than serum or
plasma results.

To compare lab results with meter
results, divide the lab result by
1.12.

Example:

Assume that your lab result (plas-
ma or serum) is 224 mg/dL.

224 divided by 1.12 = 200 (the lab
value is now approximately equal
to a whole blood result).

Compare this figure to your meter
result. The two should be within
20% of each other.

You may still have a variation from
the result because blood glucose lev-
els can change significantly over short
periods, especially if you have recent-
ly eaten, exercised, taken medication,
or experienced stress.