Lifescan Pocket Scan Blood Glucose Monitor System User Manual

Page 86

Advertising
background image

MESSAGE

PROBLEM

WHAT TO DO

not

1. The blood or control

1. Repeat the test with a

enough

solution sample was too

new test strip and a large,

b l o o d

small or smeared.

shiny drop of blood or

retest

control solution.

2. The test strip was not

2. Repeat the test with a new

inserted far enough into

test strip, pushing the test

the test strip holder.

strip all the way into the
test strip holder.

3. Your hand or an object

3. Repeat the test. Keep the

covered the test area while

meter test area clear.

the meter was turned on.

not ok

1. If appearing when you

1. Call LifeScan

turn your meter on, your

Customer Services at

meter may have an elec-

1 800 227-8862.

tronic problem.

2. If appearing at the end of a

2. Repeat the test with a new

test, your meter may have

test strip; try not to move

been moved while applying

the meter.

blood or during the test
process.

79

MESSA

GES

Advertising