Installation

 

5 Installation 

5.1 Positioning 

Sensor  location  depends  on  the  application.  Before  deciding  on  a 
location,  you  need  to  be  aware  of  the  ambient  temperature  of  the 
location,  the  atmospheric  quality  of  the  location,  and  the  possible 
electromagnetic interference in that location, according to the sections 
described above. If you plan to use air purging, you need to have an 
air  connection  available.  Wiring  and  conduit  runs  must  be 
considered, including computer wiring and connections, if used. 

5.1.1 Distance to Object 

The  desired  spot  size  on  the  target  will  determine  the  maximum 
measurement  distance.  To  avoid  erroneous  readings  the  target  spot 
size  must  completely  fill  the  entire  field  of  view  of  the  sensor. 
Consequently,  the  sensor  must  be  positioned  so  the  field  of  view  is 
the  same  as  or  smaller  than  the  desired  target  size.  For  a  list 
indicating  the  available  optics,  see  section  3.2 

Optical  Specifications

 

on page 6. 

The  actual  spot  size  for  any  distance  can  be  calculated  by  using  the 
following  formula.  Divide  the  distance  D  by  your  model’s  D:S 
number.  For  example,  for  a  unit  with  D:S  =  10:1,  if  the  sensor  is 
400 mm (15.7 in.) from the target, divide 400 by 10 (15.7 by 10), which 
gives you a target spot size of approximately 40 mm (1.57 in.). 