Thermocouple output, 3 thermocouple output, Installation – RayTek 54301 User Manual

Page 30: 22 mid

Advertising
background image

Installation

 

5.3.3 Thermocouple Output 

If  you  are  using  a  J‐  or  K‐  thermocouple  you  must  install  a 
compensation  cable.  The  cable  is  available  as  an  accessory 
(XXXCI1CB25  for  Type  J,  XXXCI2CB25  for  Type  K)  with  a  cable 
length of 7.5 m (24.6 ft.) 
Connect the wires according to the following table: 

J-Thermocouple

Power Supply

+

white

red-white

+

red-yellow

yellow

Table 1: Wiring the Thermocouple J Compensation Cable 

K-Thermocouple

Power Supply

+

yellow

red-yellow

+

red-white

white

Table 2: Wiring the Thermocouple K Compensation Cable 

Power +

red-yellow

Power –

yellow

TC J –

red-white

TC J +

white

Figure 12: Wiring the Thermocouple J Compensation Cable 

22 