Accessories

 

8.5 Air Cooling System 

The  sensing  head  can  operate  in  ambient  temperatures  up  to  200°C 
(392°F)  with  the  air‐cooling  system.  The  air‐cooling  system  comes 
with a T‐adapter including 0.8 m (7.5 in) air hose and insulation. The 
T‐adapter  allows  the  air‐cooling  hose  to  be  installed  without 
interrupting the connections to the box. 
The air‐cooling jacket may be combined with the right angle mirror. 

Air cooling (max. 35°C / 95°F)

Cable

Electronic

Housing

T-

Adapter

Air Hose

max. ambient 50°C (122°F)

Sensing Head

max. ambient 200°C (392°F)

Figure 35: Air Cooling System 

 

T-Adapter

Hose to
sensing head

Cable to electronic
housing

Fitting free for air
connection

Figure 36: Connecting the T‐Adapter (XXXMIACCJ) 