6.4.2 Peak Hold 

The output signal follows the object temperature until a maximum is 
found. Once  the  hold time is exceeded the output  signal, tracks and 
output the actual object temperature and the algorithm will start over 
again. The range for the hold time is 0.1 to 998.9 s.

 

Temp

output temperature
object temperature

hold time

hold time

Time

Figure 26: Peak Hold 

A defined hold time of 999 s (symbol “∞” in the display) will put the 
device into continuous peak detection mode. 
A  low  level  input  (GND)  at  external  input  FTC3  will  promptly 
interrupt the hold time and will start the maximum detection again. 