11.4.3 Post Processing 

The following parameters can be set to determine the post processing 
mode, see section 6.4 

Post Processing

 on page 38. 

P=5 

peak hold, hold time: 5 s 

F=12.5 

valley hold, hold time: 12.5 s 

G=10 

averaging, average time (90%): 10 s 

XY=3 

advanced peak hold, hysteresis: 3 K 

XY=‐2 

advanced valley hold, hysteresis: 2 K 

Advanced Peak/Valley Hold with Averaging: 

C=250 

threshold: 250°C 

AA=15 

averaging time (90%): 15 s 

11.5 Dynamic Data 

All temperature related information is calculated 128 times a second. 
To request the dynamic data, following commands are available: 

?T 

target temperature 

?I 

internal temperature of the sensing head 

?XJ 

internal temperature of electronics housing 

?Q 

energy value of the infra‐red temperature 

?XT 

trigger set point (active/inactive) for the FTC3 input 

To  check  for  resets  (e.g.  power  shut  down)  use  the  command  XI. 
Notice, after a reset the unit is new initialized. 

?XI 

asks for the reset status 

!XI0 

no reset occurred 

!XI1 

a reset occurred, new initialization of the unit 

XI=0 

sets the reset status back to 0 

MID 

71 