Programming guide, Mid 77 – RayTek 54301 User Manual

Page 85

Advertising
background image

Programming Guide 

Description

Char Format

P B S Legal values

Factory

default

LCD

Source: emissivity /

setpoint for alarm output

ES X

* * 1=constant

number

(E=0.950)

E=external

analogous input

FTC1

D= E/XS digital

selected FTC1-3

1

Presel. emissivity value

EV n.nnn

*

* 0.100 - 1.100

Valley hold time(4)

F nnn.n

* * * 0.000 - 998.9 s

(999 = infinite)

000.0 s

V

Average time

G nnn.n

* * * 000.0 – 999 s

000.0 s

A

Top of mA/mV range

H nnn.n

* * * (1)

(7)

H

Sensor / head ambient

I

nnn.n

* *

in current scale

(°C/°F)

Switch panel lock

J X

*

* L=locked

U=unlocked

unlocked

Alarm output control

K N

*

* 0=off

1=on

2=Target.; norm.

open

3=norm. closed

4=Head; normal

open

5=norm. closed

7=sensor / head

ambient

?

Bottom of mA/mV range

L nnn.n

* * * (1)

(8)

L

Output voltage

O n.nnn

*

* 0.000 – 5.000

voltage in V

6=controlled by

unit

6

Output current

O nn.nn

*

* 0.00 – 20.00

current in mA

21=over range

60=controlled by

unit

60 L

Peak hold time (4)

P nnn.n

* * * 000.0 998.9 s

(999 = infinite)

000.0 s

P

Power Q

nnnn

*

*

MID 

77 