Installation

 

5.4.1 Emissivity Setting (analog controlled) 

The input FTC1 can be configured to accept an analog voltage signal 
(0  to  5  VDC)  to  provide  real  time  emissivity  setting.  The  following 
table shows the relationship between input voltage and emissivity. 

U in V

0.0 0.5 … 4.5 5.0

Emissivity

0.1 0.2 … 1.0 1.1

Table 4: Ratio between Analog Input Voltage and Emissivity 

Example: 
The process requires the setting of emissivity: 

• for product 1:  0.90 
• for product 2:  0.40 

Following  the  scheme  below,  the  operator  needs  only  to  switch  to 
position “product 1” or “product 2”. 

 

+ 5 VDC

“product 1”

R1 = 200 Ω

4.0 V (ε=0.9)

To the input FTC1

of the sensor

R2 = 500 Ω

1.5 V (ε=0.4)

“product 2”

R3 = 300 Ω

Figure 14: Adjustment of Emissivity at Input FTC1 (Example) 