A Trimble Standard Interface Protocol

82 Trimble ThunderBolt E GPS Disciplined Clock User Guide

5

Not used

6

Disciplining Disabled

3

Self-Survey Progress

UINT8

0-100%

4-7

Holdover Duration

UINT32

0

Seconds

8-9

Critical Alarms

UINT16

Bit field

Bit 4: DAC at rail

10-11

Minor Alarms

UINT16

Bit field

Bit 0: DAC near rail

Bit 1: Antenna open

Bit 2: Antenna shorted

Bit 3: Not tracking satellites

Bit 4: Not disciplining oscillator

Bit 5: Survey-in progress

Bit 6: No stored position

Bit 7: Leap second pending

Bit 8: In test mode

Bit 9: Position is questionable

Bit 10: Not used

Bit 11: Almanac not complete

Bit 12: PPS not generated

12

GPS Decoding Status

UINT8

0x00

Doing fixes

0x01

Don't have GPS time

0x03

PDOP is too high

0x08

No usable sats

0x09

Only 1 usable sat

0x0A

Only 2 usable sats

0x0B

Only 3 usable sats

0x0C

The chosen sat is unusable

0x10

TRAIM rejected the fix

13

Disciplining Activity

UINT8

0

Phase locking

1

Oscillator warm-up

2

Frequency locking

3

Placing PPS

4

Initializing loop filter

5

Compensating OCXO (holdover)

6

Inactive

7

Not used

8

Recovery mode

9

Calibration/control voltage

14

Spare Status 1

UINT8

0

15

Spare Status 2

UINT8

0

16-19

PPS Offset

Single

ns

20-23

Clock Offset

Single

ppb

24-27

DAC Value

UINT32

Byte

Item

Type

Value

Description