Cursor menu, Moving the cursors – Tektronix Waveform Monitors WFM700 Series User Manual

Page 33

Advertising
background image

Menus

2- 18

WFM700 Series Waveform Monitors User Manual

Cursor Menu

If the instrument is in a mode that supports measurement cursors (Waveform,
Eye, or Jitter modes), the Cursor menu shown in Figure 2--9 is displayed when
you press the CURSOR button. The cursor 1 and 2 settings, and the difference
between the two, are shown in the readout in the upper right part of the screen.

SWEEP

GAIN

CURSOR

LINE SEL

INPUT

Voltage | Time | Voltage + Time

Cursor 1 < . . . >

Cursor to Center

V+T Control

Cursor 2 < . . . >

Cursor Menu

Track Cursors

Set Units & 100%

Percent Cursors Menu

Time: sec | 1/T

Voltage: mV | %

Done

Set 100%

Figure 2- 9: Cursor menu diagram

The selected (active) cursors are red and the inactive cursors are blue. To move

the active cursor:

1.

Touch the Voltage/Time soft key to select the cursor type: voltage, time, or
both voltage and time. If you select both, use the V+T Control soft key to
toggle between adjusting voltage cursors and adjusting time cursors.

2.

Touch the corresponding soft key to select the active cursor: Cursor 1,
Cursor 2, or Track Cursors. (Use Track Cursors if you want to adjust the
position of either the voltage or the time cursors as a pair.) If Track Cursors
is not selected, you can use the SELECT button to toggle between Cursor 1
and Cursor 2 as the active cursor.

3.

Turn the general purpose knob to move the active cursor(s). The correspond-
ing readouts and soft key displays update as you move the cursors.

4.

You can touch the Cursor to Center soft key to move the selected cursor to
the center of the display.

NOTE. Mag or Gain settings other than X1 may cause the cursors to be off
screen.

Moving the Cursors