Reprocessing the mask between patients

The mask should be reprocessed when used between patients. Cleaning,
disinfection and sterilisation instructions are available from the ResMed website,

www.resmed.com/masks/sterilization. If you do not have internet access,
please contact your ResMed representative.

Technical specifications

Pressure-flow curve The mask contains passive venting to protect against

rebreathing. As a result of manufacturing variations, the vent flow rate may vary.

Vent flo

w rat

e (L/min)

0

20

40

60

80

2

4

6

8

10

12

14

16

18

20

22

24

26

28

30

32

34

36

38

40

Quattro Air/ Quattro Air for Her

Quattro

Quattro FX/ Quattro FX for Her

Liberty

Mask pressure (cm H

2

O)

Quattro Air/

Quattro Air

for Her

Quattro FX/

Quattro FX

for Her

Quattro

Liberty

Pressure (cm H

2

O) Flow (L/Min)

3

19

4

22

22

22

22

8

32

32

32

32

Quattro Air/

Quattro Air

for Her

Quattro FX/

Quattro FX

for Her

Quattro

Liberty

Pressure (cm H

2

O) Flow (L/Min)

10

37

37

37

37

12

41

41

41

41

16

48

48

48

48

20

54

54

54

54

24

60

60

60

28

66

66

66

30

69

69

69

32

72

72

36

77

77

40

82

82

Deadspace: Empty volume of mask to swivel. The deadspace of the mask varies
according to cushion size.

222 mL
(Large)

243 mL
(Large)

240 mL
(Large)

151 mL
(Large)*

* Empty volume of mask to elbow.
Therapy pressure (cm H

2

O)

3 - 40

4 - 30

4 - 40

4 - 20

Resistance: Drop in pressure measured (nominal) (cm H

2

O)

at 50 L/min

0.2

0.1

0.1

0.1

at 100 L/min

0.9

0.5

0.4

0.5

Inspiratory and Expiratory Resistance (cm H

2

O)

Inspiration

0.6

0.7

0.8

0.7

Expiration

1.0

0.7

0.8

0.8