Storage, Disposal, Warnings – ResMed Mirage Liberty User Manual

Page 25

Advertising
background image

25

Quattro Air/

Quattro Air

for Her

Quattro FX/

Quattro FX

for Her

Quattro

Liberty

Anti asphyxia valve (AAV) open-to-atmosphere pressure (cm H

2

O)

≤ 1.6

<= 1.8

<= 1.8

Anti asphyxia valve (AAV) closed-to-atmosphere pressure (cm H

2

O)

≤ 1.8

<= 1.8

<= 1.8

Sound - DECLARED DUAL-NUMBER NOISE EMISSION VALUES in accordance with ISO
4871. The A-weighted sound power level and the A-weighted sound pressure level of
the mask, at a distance of 1M, with uncertainty of 3 dBA, are shown.
Power Level

28 dBA

27 dBA

26 dBA

24 dBA

Pressure level
at 1m

20 dBA

19 dBA

18 dBA

16 dBA

Gross Dimensions - Mask fully assembled - no headgear. Largest variant only shown.

Height 172 mm

140 mm

206 mm

100 mm

Width 103 mm

176 mm

109 mm

110 mm

Depth 107 mm

144 mm

108 mm

100 mm

Mask setting options: Select A if available, otherwise select B.

A Full Face

Mir Full

Mir Full

Mir Full

B Mir Full

Full Face

Full Face

Full Face

Environmental conditions – all masks:

Operating temperature: +5°C to +40°C

Operating humidity: 15% to 95% non-condensing

Storage and transport temperature: –20°C to +60°C

Storage and transport humidity:

up to 95% non-condensing

Storage

Ensure that the mask is thoroughly clean and dry before storing it for any length
of time. Store the mask in a dry place out of direct sunlight.

(Quattro FX/Quattro FX for Her only) To maintain optimum care of your mask,
ensure the spring frame support is fitted when travelling.

Disposal

The mask does not contain any hazardous substances and may be disposed of
with your normal household refuse.
The following symbols may appear on the mask or packaging.

Caution, consult accompanying documents;

LOT

Batch Code;

REF

Catalogue

Number;

European Authorised Representative;

Temperature

limitation;

Humidity limitation;

Not made with natural rubber latex;

Manufacturer;

Keep away from rain;

This way up;  Fragile, handle

with care.

Warnings

• The mask should only be used with CPAP or bilevel devices recommended by a

physician or respiratory therapist.

• As with all masks, some rebreathing may occur at low CPAP pressures.

• The vent holes must be kept clear.

• The mask should not be used unless the device is turned on. Once the mask is

fitted, ensure the device is blowing air.

• Follow all precautions when using supplemental oxygen.

• Oxygen flow must be turned off when the CPAP or bilevel device is not

operating so that unused oxygen does not accumulate within the device
enclosure and create a risk of fire.

• At a fixed flow rate of supplemental oxygen flow, the inhaled oxygen

concentration varies depending on the pressure settings, user breathing
pattern, mask, point of application and leak rate.

• If any visible deterioration of a mask component is apparent (cracking, tears,

etc), the mask component should be discarded and replaced.

• Always follow cleaning instructions and use mild soap. Some cleaning products

may damage the mask, its parts and their function, or leave harmful residual
vapours that could be inhaled if not rinsed thoroughly.