Storing Results

The Accu-Chek Advantage meter automatically stores 100
glucose results, letting you review them in order from the
most recent to the oldest. If you have set the time/date fea-
ture, the time and date of the results are also displayed.
There is a situation that can occur no more than once
every seven days during a ten minute period, which can
cause an individual result to be stored incorrectly in the
memory of the Accu-Chek Advantage meter. If time and
date are set on your meter, this sitation will only occur
when your meter clock reads 12:00 (midnight) - 12:10 a.m.
If time and date are not set on your meter, this situation
can occur once every seven days at any time of day. Do
not change your therapy based on one individual result
recorded in memory.

Other Information

83

10. To perform another

control check using
Level 2 glucose con-
trol solution, return
to step 4 and follow
the same steps
again. Turn meter
OFF when you are
done with your qual-
ity control check.

Quality Control Testing

82

IF...

the memory is full with
100 glucose results,

the memory is full and a
new result is added,

THEN...

“mem” flashes in the dis-
play when the meter is
turned on.

meter deletes the
oldest result.

Results stored in memory will not be lost when

you replace the batteries.