CAUSE

Do not use one individual result stored in memo-
ry to make therapy adjustments.

Other Informtion

89

DISPLAY

Result that appears in
memory does not match
the result seen on the
screen at the time the
test was performed or
is outside the 10-600
mg/dL reading range.

CAUSE

If a test is performed at
precisely the same
moment the Accu-Chek
Advantage meter is
checking its internal cal-
endar and clock settings,
which happens automati-
cally once every seven
days, an incorrect test
result can be stored in
the meter memory.
If time and date are set
on your meter, this situa-
tion will only occur once
every seven days when
your meter clock reads
from 12:00 (midnight) -
12:10 am. If time and
date are not set on your
meter, this situation can
occur once every seven
days at any time of day.

88

Troubleshooting Guide

The following are display messages and error messages
that may appear on the meter. If you are unable to resolve
a concern, please call Accu-Chek Customer Care

SM

at

1-800-858-8072, 24 hours a day, 365 days a year.