Chapter 5: Maintenance and Troubleshooting

77

ACCU-CHEK Customer Care Service Center 1-800-858-8072

The test drum 90-day use-by period

has expired. If the correct time and

date were set on the meter and the

drum door has not been opened

since the installation of the test

drum, discard the old test drum and

insert a new test drum.

The drum door was opened and

closed with a partly used test

drum inside, or a partly used test

drum was inserted.
If there was a partly used test

drum in the meter and it took

longer than two minutes to

replace the batteries, the year,

date, and time have to be reset.

When the year, date, and time

are reset, the meter does not

know how long the partly used

test drum has been in the meter.
You should run a control test

to verify that the test drum is

useable. See Chapter 2, “Why

Run Control Tests?”