Chapter 5: Maintenance and Troubleshooting

81

ACCU-CHEK Customer Care Service Center 1-800-858-8072

• The test strip is stuck in the test

strip slot or the test drum does

not rotate. Pull the test strip out

of the meter and discard it. Then,

take the test drum out and reinsert

it. Remember the 90-day use-by

counter in the meter is no longer

correct. The drum symbol flashes

until you install a new test drum.

• The meter was turned off and on

again, but the test strip is still in

the meter. Hold the meter straight

up and down and let the test strip

drop out, or take it out manually,

even if it takes a small amount of

force.

The test strip was bent or removed

during testing. Remember to slide

and release

, at the end of the

test, to turn the meter off and allow

the test strip to release. Slide and

release

to turn on the meter

again and a new test strip

advances.