Hold button (ma95z power unit), Firm/soft buttons, Automatic wireless fowler – Activeforever Invacare microAIR Lateral Alternating Pressure Mattress & Compressor (microAIR Mattress with Compressor 36 x 80 x 15) User Manual

Page 16: Manual operation of the fowler, About automatic wireless mode

Advertising
background image

SECTION 2—OPERATION

MA90 and MA95 Series Mattresses

16

Part No 1148139

3. Use the Select button to move the pointer to the times or attributes available for the 

selected function.

4. Use the Firmer or Softer buttons to change the times or attributes.

Hold Button (MA95Z Power Unit)

1. Press the Hold button (

) to pause the mattress in the current position. The mattress 

will remain in this position until the Hold button is pressed again.

Firm/Soft Buttons

1. Select comfort pressure settings by pressing Firm (

) or Soft (

) buttons.

• Soft Button ‐ Pressing this button reduces the pressure in the mattress.

• Firm Button ‐ Pressing this button increases the pressure in the mattress.

NOTE: The patient comfort pressure ranges from Soft (6±4 mmHg) to Firm (32±6 mmHg). The 
Comfort Control LED displays the patient comfort pressure levels and provides a guide to the 
caregiver to set approximate comfort pressure level depending on the patient weight. If the patient’s 
weight to height ratio is above average, increase the pressure setting by approximately 20%. 
MA90Z models display levels from 0 to 9 and MA95Z models display the mmHg from 6 to 32. 

Automatic Wireless Fowler

Manual Operation of the Fowler

1. Press the Select button (

) to move the pointer on the LCD to Function.

2. Use the Firm and Soft buttons until Fowler displays.

NOTE: When this mode is activated, the control unit increases the pressure in the mattress to 
prevent the patient from bottoming out.

About Automatic Wireless Mode

When the bed articulates to 45°, the transmitter in the head of the mattress signals the 
control unit to increase the pressure in the mattress by 80% to prevent the patient from 
bottoming out.

PULSE

Pulse mode. The pulse time indicates the time between
pulses. For example, 5 minutes is selected. In this
example, all of the air cushions in the mattress will
deflate for 4 seconds and inflate for 5 minutes (the
selected time). The duration can be set from 1 to 99
minutes.

FOWLER, UPRIGHT

Fowler Mode. The pressure will increase to prevent the
patient from bottoming out.

PULSE N STATIC

10 minute therapy duration until turned off.

FUNCTION

DESCRIPTION