Activeforever Invacare microAIR Lateral Alternating Pressure Mattress & Compressor (microAIR Mattress with Compressor 36 x 80 x 15) User Manual

Page 17

Advertising
background image

SECTION 2—OPERATION

Part No 1148139

17

MA90 and MA95 Series Mattresses

Programming the Automatic Wireless Fowler Transmitter

1. Ensure the transmitter is inside the mattress pocket located on the right side of the 

mattress at the head end.

2. Ensure the mattress is in a flat position.

3. Ensure the control unit is plugged in, and the LCD displays STAND BY.

4. Perform one of the following:

• MA90Z Power Unit ‐ Perform the following steps:

i. Press and hold the Turn Angle and Max Inflate buttons.

NOTE: The Fowler LED is lit and L displays in the LCD Display.

ii. Raise the bed to 45°.

NOTE: The control unit will beep and will return to Standby.

iii. Perform one of the following:

• Cancel ‐ Press the Turn Angle button to exit Fowler Transmitter Setup 

without linking to a transmitter.

• Save ‐ Wait 60 seconds without pressing a key. 

• MA95Z Power Unit ‐ Perform the following steps:

i. Press and hold the Hold and Max Inflate buttons.

ii. AUTO FOWLER LEARN MODE will display.

iii. PLACE BED IN FLAT POSITION will display.

iv. CONTINUE ‐> SELECT will display.

v. Press the Select button.

vi. PLACE BED IN FOWLER/UPRIGHT POSITION will display.

vii. TRANSMITTER DETECTED will display.

viii.The control unit will beep.

ix. STAND BY will display.

Advertising