Mode button (ma90z power unit), Max inflate button (ma95z power unit) – Activeforever Invacare microAIR Lateral Alternating Pressure Mattress & Compressor (microAIR Mattress with Compressor 36 x 80 x 15) User Manual

Page 18

Advertising
background image

SECTION 2—OPERATION

MA90 and MA95 Series Mattresses

18

Part No 1148139

Mode Button (MA90Z Power Unit)

1. Press the Mode button (

) to select the Max Inflate Mode or the Low Air Loss mode.

• Max Inflate Mode ‐ In this mode, the mattress inflates rapidly to maximum 

firmness (pressurized to 35 ± 6 mmHg). A series of beeps sound every three 
minutes as a reminder that the Max Inflate mode is active. After 30 minutes, the 
Max Inflate mode deactivates and the control unit defaults to the previous setting. 

NOTE: It takes 5‐10 minutes for the MA90Z mattress to inflate fully (inflation time depends 
on size of mattress).

NOTE: Max Inflate mode can be manually disengaged by pressing the Max Inflate button. 
This will deactivate the Max Inflate LED.

NOTE: It is recommended that Max Inflate setting be used during patient ingress/egress, 
patient wound care, patient turning or patient cleaning.

• Low Air Loss Mode ‐ In this mode, the mattress goes into on‐demand low air loss 

relief mode.

Max Inflate Button (MA95Z Power Unit)

1. Press the Max Inflate button (

) to select the max inflate mode and inflate the 

mattress rapidly to maximum firmness (pressurized to 35 ± 6 mmHg). A series of 
beeps sound every three minutes as a reminder that the Max Inflate mode is active. 
After 15 minutes, the Max Inflate mode deactivates and the control unit defaults to the 
previous setting. 

NOTE: It takes 25‐60 seconds for the MA95Z mattress to inflate fully (inflation time depends on 
size of mattress).

NOTE: It is recommended that Max Inflate setting be used during patient ingress/egress, patient 
wound care, patient turning or patient cleaning.

2. Perform one of the following:

• MA90Z Power Unit ‐ Press the Max Inflate button again to manually disengage 

Max Inflate mode. 

• MA95Z Power Unit ‐ Select a different setting to exit Max Inflate mode, or press the 

Max Inflate button again. 

NOTE: This will deactivate the Max Inflate LED.

Advertising