MICRO-LINE

Reference Manual

SETUP

11

Testing the Scales

Follow these steps to confirm that your scales have been installed proper-
ly. This test will confirm the scale’s electrical operation, and will also
check the installation integrity.

Locate a magnetic base on the machine and set the dial indicator
on the scale’s reading head casting. Zero the readout and the
indicator.

Move the axis through the full travel and return the dial to “0.”
The readout should also read 0 (±.0005” for 10µm scales;
±0.0002” for 5µm scales). If it doesn’t, then the scale cable may
be loose, or the scale, mounting bracket, or reading head may
need to be tightened or realigned.

Repeat these steps for each scale.