Veris Technologies pH Manager - Operating Instructions User Manual

Page 59

Advertising
background image

Pub.# OM17-MSP-pH

6-14

3. Testing continuity from the circuit board through the communication cable

a) Remove the steel cover and unscrew the plastic lid from the external Controller.
b) Look for the brown Molex 3-pin connector circled in Figure 4. The color sequence

(left->right) should be white-green-black. CAREFULLY wiggle each wire to check for
a loose connection.

c) Check continuity between the following pins:












Figure 3.4 Brown Molex connector on circuit board (POWER MUST BE DISCONNECTED
BEFORE TESTING)

d) Check continuity from brown Molex connector to 9 pin serial connector; this tests
continuity from the circuit board to the DataLogger or PC. (Figure 5)




















Figure 3.5 Testing continuity from Molex connector to serial connector

Molex Connector 4-pin connector

9 pin serial connector

White

2

3

Green

1

2

Black

3

5

Advertising