3 measurement of liquids, Flat vessel top – VEGA VEGASON 54P…56P Profibus PA User Manual
Page 17
VEGASON 54P … 56P
17
< 100 mm
It is also possible to mount version C sensors
in a 1" thread.
2.3 Measurement of liquids
Flat vessel top
On flat vessels, mounting is usually done on
a very short DIN socket piece. The lower
side of the flange is the reference plane. The
transducer should protrude out of the flange
pipe.
VEGASON 55 in flange version on short DIN socket
piece
Reference plane
Mounting and installation
Min. distance
Type 54: 1 m
Type 55: 0.8 m
< 400 mm
VEGASON 56 in flange version on short DIN socket
piece
Reference plane
Min. distance
1.8 m
A mounting location directly on the vessel top
is ideal. A round opening in the vessel top is
sufficient to fasten the sensor with the flange
or version B and C with the swivelling holder.
Flange version and swivelling holder on flat vessel
top
Reference
plane
Reference
plane
< 60 mm
Mounting of the transducer in a 1" thread
Reference plane
Min. meas. distance
1.8 m
Min. meas. dis-
tance 1.4 m
Swivelling
holder