Parameter adjustment – VEGA VEGAMET 515 User Manual

Page 48

Advertising
background image

48

VEGAMET 515

1)

1)

Parameter adjustment

7.11

Special function: Manual and automatic correction

Manual offset correction

(only with hydrostatic pressure transmitters)
This correction can be carried out to com-
pensate changes of the sensor values due to
ageing. In order to do this, the sensor must
be in an unpressurized condition (empty
vessel).

Param.
TAG-
Nr. 1

TAG -
Nr. 1
%
xx,x

Adjust-
ment

Signal
condit-
ioning

Out-
puts

Simu-
lation

Spe-
cal
funct.

Reset
Level

Manu-
al
corr.

Offset
correc-
tion

Offset
correc-
tion
OK?

Correc-
tion
Now !
OK ?

Correc-
tion
Now !
OK ?

Real
value
corr.

Real
value
corr.
OK?

Correc-
tion
%

0,0

Manual real value correction

In the menu item "Correction % 0.0“, you can
enter a percentage value which corresponds
to the actual existing level.

Automatic correction:

1)

This menu item only appears if in the measurement
loop configuration a correction mode was set. Then
you enter here the percentage value to which the
automatic correction responds.

Reset

OK ?

Reset
Now!

OK ?

Simu-
lation
%

XX,X

Simu-
lation
Now!
OK?

DK
corr.
at %

50,0

Density
corr.
at %

50,0

Real
value
corr.
at %

0

1)