VEGA VEGAMET 515 User Manual

Page 59

Advertising
background image

VEGAMET 515

59

Error code

Meaning

Diagnosis

E061

Span of the voltage output values too small. The values for the voltage output
are too close together or are identical. Please check the adjustment of the
voltage output.

E063

Wrong sensor characteristics values. Please check the sensor characteristics
values

E102

A reset of the configuration was carried out on this VEGAMET via the key-
board.

E110

Switching hysteresis too small. The switching points are too close together or
are identical.

E111

Relay switching points reversed. The relay switching points are reversed in
relation to the selected relay mode.

Advertising