VEGA VEGAPULS 62 (≥ 2.0.0 - ≤ 3.8) 4 … 20 mA_HART four-wire User Manual

Page 19

Advertising
background image

d

h

max.

d

1½"

50 mm/2"

80 mm/3"

100 mm/4"

150 mm/6"

200 mm

250 mm

300 mm

500 mm

800 mm

h

max.

Fig. 11: Deviating socket dimensions

Tip:
VEGAPULS

62

is optionally also available with antenna extension.

H

ence the antenna length can be selected such that the antenna end

protrudes 10 mm (0.4 in) out of the socket.

A

lign the sensor in liquids as vertical as possible to the product surface

to achieve optimum measuring results.

Fig. 12: Alignment in liquids

T

he mounting location of the radar sensor should be a place where no

other equipment or fixtures cross the path of the microwave signals.

V

essel installations such as, for example, ladders, limit switches,

heating spirals, struts, etc. can cause false echoes that get super-
imposed on the useful echo. Make sure when planning your measuring
site that the radar sensor has a "clear view" to the measured product.

I

n case of existing vessel installations, a false echo storage should be

carried out during setup.

I

f large vessel installations such as struts or supports cause false

echoes, these can be attenuated through supplementary measures.
S

mall, inclined sheet metal baffles above the installations scatter the

radar signals and prevent direct interfering reflections.

Sensor orientation

Vessel installations

VEGAPULS

62 • 4 … 20

mA/HART four-wire