Fig. 17: Polarisation level with flange version, view from top on the sensor and
bypass tube. The sensor housing is not shown.

1

Position of the polarisation level

2

Flange hole

3

Upper tube connection

4

Polarisation marking

T

he short examples are meant to give only introductory information on

the flow measurement. Planning information is available from flume
manufacturers and in special literature.

Meas. wall with rectangular flume

I

n general, the following points must be observed:

l

I

nstallation of the sensor on the headwater side

l

I

nstallation in the center of the flume and vertical to the liquid

surface

l

D

istance to the overfall orifice

l

D

istance of orifice opening above ground

l

M

in. distance of the orifice opening to bottom water

l

M

in. distance of the sensor to max. storage level

Flow measurement

VEGAPULS

62 • 4 … 20

mA/HART four-wire

23