VEGA VEGABAR 65 Foundation Fieldbus User Manual
Page 58
-
Double chamber housing
l
1 x IP 68 cable gland M20 x 1.5; 1 x blind
stopper M20 x 1.5; plug M12 x 1 for VEGADIS
61 (optional)
or:
l
1 x closing cap ½ NPT, 1 x blind stopper
½ NPT, plug M12 x 1 for VEGADIS 61
(optional)
Connection cable
-
Configuration
four wires, one suspension cable, one breather
capillary, screen braiding, metal foil, mantle
-
Wire cross-section
0.5 mm² (AWG 20)
-
Wire resistance
< 0.036 Ω/m (0.011 Ω/ft)
-
Tensile strength
> 1200 N (270 pounds force)
-
Standard length
5 m (16.4 ft)
-
Max. length
1000 m (3281 ft)
-
Min. bending radius at 25 °C/77 °F
25 mm (0.985 in)
-
Diameter approx.
8 mm (0.315 in)
-
Colour - standard PE
Black
-
Colour - standard PUR
Blue
-
Colour - Ex-version
Blue
Electromechanical data - version IP 68
Cable entry/plug
15)
-
External housing
l
1 x cable gland M20 x 1.5 (cable: ø 5 … 9 mm),
1 x blind stopper M20 x 1.5
or:
l
1 x plug (depending on the version), 1 x blind
stopper M20 x 1.5
Spring-loaded terminals for wire cross-sec-
tion up to
2.5 mm² (AWG 14)
Connection cable between IP 68 instrument and external housing:
-
Configuration
four wires, one suspension cable, one breather
capillary, screen braiding, metal foil, mantle
-
Wire cross-section
0.5 mm² (AWG 20)
-
Wire resistance
< 0.036 Ω/m (0.011 Ω/ft)
-
Standard length
5 m (16.40 ft)
-
Max. length
180 m (590.5 ft)
-
Min. bending radius at 25 °C/77 °F
25 mm (0.985 in)
-
Diameter approx.
8 mm (0.315 in)
-
Colour - standard PE
Black
15)
Depending on the version M12 x 1, according to DIN 43650, Harting, 7/
8" FF.
58
VEGABAR
65 • Foundation Fieldbus
10 Supplement
28260
-EN
-090522