Double chamber housing

l

1 x IP 68 cable gland M20 x 1.5; 1 x blind
stopper M20 x 1.5; plug M12 x 1 for VEGADIS
61 (optional)

or:

l

1 x closing cap ½ NPT, 1 x blind stopper
½ NPT, plug M12 x 1 for VEGADIS 61
(optional)

Connection cable

-

Configuration

four wires, one suspension cable, one breather
capillary, screen braiding, metal foil, mantle

-

Wire cross-section

0.5 mm² (AWG 20)

-

Wire resistance

< 0.036 Ω/m (0.011 Ω/ft)

-

Tensile strength

> 1200 N (270 pounds force)

-

Standard length

5 m (16.4 ft)

-

Max. length

1000 m (3281 ft)

-

Min. bending radius at 25 °C/77 °F

25 mm (0.985 in)

-

Diameter approx.

8 mm (0.315 in)

-

Colour - standard PE

Black

-

Colour - standard PUR

Blue

-

Colour - Ex-version

Blue

Electromechanical data - version IP 68

Cable entry/plug

15)

-

External housing

l

1 x cable gland M20 x 1.5 (cable: ø 5 … 9 mm),
1 x blind stopper M20 x 1.5

or:

l

1 x plug (depending on the version), 1 x blind
stopper M20 x 1.5

Spring-loaded terminals for wire cross-sec-
tion up to

2.5 mm² (AWG 14)

Connection cable between IP 68 instrument and external housing:

-

Configuration

four wires, one suspension cable, one breather
capillary, screen braiding, metal foil, mantle

-

Wire cross-section

0.5 mm² (AWG 20)

-

Wire resistance

< 0.036 Ω/m (0.011 Ω/ft)

-

Standard length

5 m (16.40 ft)

-

Max. length

180 m (590.5 ft)

-

Min. bending radius at 25 °C/77 °F

25 mm (0.985 in)

-

Diameter approx.

8 mm (0.315 in)

-

Colour - standard PE

Black

15)

Depending on the version M12 x 1, according to DIN 43650, Harting, 7/
8" FF.

58

VEGABAR

65 • Foundation Fieldbus