VEGA VEGABAR 65 Foundation Fieldbus User Manual

Page 59

Advertising
background image

-

Colour - standard PUR

Blue

-

Colour - Ex-version

Blue

Indicating and adjustment module

Voltage supply and data transmission

through the sensor

Indication

LC display in dot matrix

Adjustment elements

4 keys

Protection rating

-

unassembled

IP

20

-

mounted into the sensor without cover

IP

40

Materials

-

Housing

ABS

-

Inspection window

Polyester foil

Power supply

Operating voltage

-

Non-Ex instrument

9 … 32 V DC

-

EEx-ia instrument

9 … 24 V DC

-

EEx-id instrument

9 … 32 V DC

Operating voltage with lighted indicating and adjustment module

-

Non-Ex instrument

12 … 32 V DC

-

EEx-ia instrument

12 … 24 V DC

-

EEx-id instrument

12 … 32 V DC

Power supply by/max. number of sensors

-

H

1 power supply

max. 32 (max. 10 with Ex)

Electrical protective measures

Protection rating

-

Housing, standard

IP

66/IP 67

16)

-

Aluminium and stainless housing (op-
tionally available)

IP

68 (1 bar)

17)

-

Process component in IP 68 version

IP

68

-

External housing

IP

65

Overvoltage category

III

Protection class

II

16)

Instruments with gauge pressure measuring ranges cannot detect the am-
bient pressure when submerged, e.g. in water. This can lead to falsification
of the measured value.

17)

Only with instruments with absolute pressure ranges.

VEGABAR

65 • Foundation Fieldbus

59

10 Supplement

28260

-

EN

-090522

Advertising