2 mounting steps, 3 mounting steps external housing – VEGA VEGABAR 64 4 … 20 mA_HART User Manual

Page 13

Advertising
background image

1

2

Fig. 4: Temperature ranges
1

Process temperature

2

Ambient temperature

4.2 Mounting steps

F

or mounting VEGABAR 64, a welded socket is required. You can find

these components in the supplementary instructions manual "Welded
socket and seals
".

S

eal the thread with teflon, hemp or a similar resistant seal material on

the process fitting thread 1½ NPT.

à Screw VEGABAR 64 into the welded socket. Tighten the hexagon

on the process fitting with a suitable wrench. Wrench size, see
chapter "Dimensions".

Warning:
T

he housing must not be used to screw the instrument in! Applying

tightening force can damage internal parts of the housing.

S

eal the flange connections according to DIN/ANSI with a suitable,

resistant seal and mount VEGABAR 64 with suitable screws.

U

se the seal suitable for the respective process fitting. You can find

the components in the supplementary instructions manual "Welded
socket and seals
".

4.3 Mounting steps external housing

1

M

ark the holes according to the following drilling template

2

D

epending on the mounting surface, fasten the wall mounting

plate with 4 screws

Welding the socket

Sealing/Screwing in
threaded versions

Sealing/Screwing in
flange versions

Sealing/Screwing in hy-
gienic fittings

Wall mounting

VEGABAR

64 • 4 … 20

mA/HART

13

4 M

ounting

27525

-

EN

-081209