P

roceed as follows:

1

E

dit the mbar value in the menu item "zero" with [OK].

Z

ero adjustment

000.0 %

P

=

+0000

.0 mbar

0000.0

mbar

2

S

et the requested mbar value with [+] and [->].

3

C

onfirm with [+] and move to span adjustment with [->].

T

he zero adjustment is finished.

Information:
T

he zero adjustment shifts the value of the span adjustment. The span,

i.e. the difference between these values, however, remains un-
changed.

Information:
F

or an adjustment with pressure, you simply enter the displayed actual

measured value. If the adjustment ranges are exceeded, the message
"Outside parameter limits" appears. The editing procedure can be
aborted with [ESC] or the displayed limit value can be accepted with
[OK]

.

P

roceed as follows:

1

E

dit the mbar value in the menu item "span" with [OK].

S

pan adjustment

100.0 %

P

=

+1000

.0 mbar

0000.0

mbar

Information:
T

he displayed pressure for 100 % corresponds to the nominal

measuring range of the sensor (in the above example 1 bar =
1000

mbar).

2

S

et the requested mbar value with [->] and [OK].

3

C

onfirm with [OK] and move to the menu overview with [ESC].

T

he span adjustment is finished.

Information:
F

or an adjustment with pressure, you simply enter the displayed actual

measured value. If the adjustment ranges are exceeded, the message
"Outside parameter limits" appears. The editing procedure can be
aborted with [ESC] or the displayed limit value can be accepted with
[OK]

.

Carrying out zero ad-
justment

Carrying out span ad-
justment

38

VEGABAR

64 • 4 … 20

mA/HART

6 S

et up with the indicating and adjustment module PLICSCOM

27525

-EN

-081209