6.5 Menu schematic
Information:
D
epending on the version and application, the highlighted menu
windows are not always available.
Basic adjustment
1
▶
B
asic adjustment
D
isplay
D
iagnostics
S
ervice
I
nfo
1.1
U
nit
U
nit of measurement
bar
▼
T
emperature unit
°C
▼
1.2
P
osition correction
Off
set
P
=
0.2
mbar
0000
mbar
1.3
M
in. adjustment
000.0 %
=
0.0 mbar
0.0
mbar
1.4
M
ax. adjustment
100.00 %
=
100.00 mbar
0.0
mbar
1.5
D
amping
1 s
1.6
L
inearisation curve
linear
▼
1.7
S
ensor-TAG
Sensor
Display
2
B
asic adjustment
▶
D
isplay
D
iagnostics
S
ervice
I
nfo
2.1
D
isplayed value
Pressure
▼
2.1
D
isplayed value
▼
Scaled
2.2
D
isplay unit
Volume
▼
l
▼
2.3
S
caling
0 % = 0.0
l
100 % = 100.0
l
2.4
L
ighting
Switched off
▼
42
VEGABAR
64 • 4 … 20
mA/HART
6 S
et up with the indicating and adjustment module PLICSCOM
27525
-EN
-081209
Advertising