4 setup, 1 display module, Adjustment – VEGA VEGABAR 41 HART User Manual

Page 17: Setup, Adjustment elements

Advertising
background image

VEGABAR 41 (HART

®

)

17

4 Setup

Electrical connection and setup can be sim-
plified by temporarily mounting the cover or
the display module on VEGABAR laterally or
downwardly displaced.

OPERATE

t

SPAN

ZERO

i

Setup

Display module

VEGABAR

8

7

6

5

4

3

2

1

4 ¼ 20 mA

DISPLAY

VEGADIS 10

12 ¼ 36 V DC E12

+ -

+

-

OK

Tighten screws lightly

4.1 Display module

Digital value
- 4 digital and decimal point
- basic function module: indicates a perma-

nent bar measurement range

- menu-driven module: allows individual

scaling of bar measurement range

Connection is made via a plug connector
according to the "Electrical connection“ sec-
tion.

4.2 Setup with module "Adjustment

of the basic functions“

Adjustment elements

Tendency
indication

Bar graph

Digital value

Adjustment system

• Select the required function with the rotary

switch.

• Change the value with the "+“ and "–“ keys.
• Reset rotary switch to OPERATE, the ad-

justed values are transferred to the EEP-
ROM-memory. They remain there even in
case of voltage drop.

Adjustment

For adjustment of zero and span, an amme-
ter must be connected to terminals 1 and 3.
The measured value is identical to the output
current.

1 Adjustment of zero

(e.g. process pressure zero or vessel
empty)
• Set rotary switch to zero
• By pushing the "+“ and "–“ keys (simulta-

neously) the current jumps directly to 4 mA
or set it to 4 mA by pushing the "+“ and "–“
keys.

Adjustment range of zero:
-20 % … +95 % of the nominal range (corre-
sponds to turn of up to +95 %)

Rotary switch: Select requested
function

Key +: Modify
value (increase

Key –: Modify
value (de-
crease)

Advertising