2 connect the pc with vegaconnect 4 – VEGA VEGABAR 74 4 … 20 mA_HART User Manual

Page 22

Advertising
background image

7

.2 Connect the PC with VEGACONNECT 4

2

4

3

1

OPEN

TW

IST

USB

LO

CK

Fig. 9: Connecting the PC via HART to the signal cable
1

VEGABAR 74

2

HART resistance 250 Ohm (optional depending on the processing)

3

Connection cable with 2 mm pins and terminals

4

Processing system/PLC/Voltage supply

N

ecessary components:

l

VEGABAR

74

l

PC

with PACTware™ and suitable VEGA DTM

l

VEGACONNECT

4

l

HART

resistance 250 Ohm (optional depending on the

processing)

l

P

ower supply unit or processing system

Note:
W

ith power supply units with integrated HART resistance

(

internal resistance approx. 250 Ohm), an additional external

resistance is not necessary. This applies, e.g. to the VEGA
instruments VEGATRENN 149A, VEGADIS 371, VEGAMET
381). A

lso usual Ex separators are most of the time equipped

with a sufficient current limitation resistor. In such cases,
VEGACONNECT

4

can be connected parallel to the 4 … 20 mA

cable.