VEGA VEGABAR 74 4 … 20 mA_HART User Manual

Page 29

Advertising
background image

Nominal range

Overload, max. pres-
sure6)

Overload, min. pressure

-0.2 … 0.2

bar/-20 … 20 kPa

30

bar/3000 kPa

-0.8

bar/-80 kPa

-0.5 … 0.5

bar/-50 … 50 kPa

35

bar/3500 kPa

-1

bar/-100 kPa

A

bsolute pressure

0 … 0.1

bar/0 … 10 kPa

15

bar/1500 kPa

0 … 1

bar/0 … 100 kPa

35

bar/3500 kPa

0 … 2.5

bar/0 … 250 kPa

50

bar/5000 kPa

0 … 5

bar/0 … 500 kPa

65

bar/6500 kPa

0 … 10

bar/0 … 1000 kPa

90

bar/9000 kPa

0 … 25

bar/0 … 2500 kPa

130

bar/13000 kPa

0 … 60

bar/0 … 6000 kPa

200

bar/20000 kPa

Reference conditions and influencing variables (similar to DIN EN 60770-1)
R

eference conditions according to DIN EN 61298-1

-

T

emperature

+

15 … +25 °C (+59 … +77 °F)

-

R

elative humidity

45 … 75 %

-

A

ir pressure

860 … 1060

mbar/86 … 106 kPa

(12.5 … 15.4

psi)

D

etermination of characteristics

L

imit point adjustment according to

IEC

61298-2

C

haracteristics

linear

R

eference installation position

upright, diaphragm points downward

I

nfluence of the installation position

<

0.2

mbar/20 Pa (0.003 psi)

Deviation determined according to the limit point method according to IEC 60770

7

)

A

pplies to digital HART interface as well as to analogue current output 4 … 20 mA.

S

pecifications refer to the set span. Turn down (TD) = nominal measuring range/set span.

D

eviation

-

T

urn down 1:1 up to 5:1

<

0.075 %

-

T

urn down up to 10:1

<

0.015 %

x TD

D

eviation with absolutely flush process fittings EV, FT

-

T

urn down 1:1 up to 5:1

<

0.05 %

-

T

urn down up to 10:1

<

0.01 %

x TD

7)

I

ncl. non-linearity, hysteresis and non-repeatability.

S

upplement

VEGABAR

74 - 4 … 20

mA/HART

29

28432

-

EN

-070718

Advertising