VEGA VEGASON 63 4 … 20 mA_HART - two-wire User Manual

Page 44

Advertising
background image

44

10 Supplement

VEGASON 63 • 4 … 20 mA/HART - two-wire

28777-EN-130417

Process conditions
Process pressure

Ʋ with compression flange

-20 … 100 kPa/-0.2 … 1 bar (-2.9 … 14.5 psi)

Ʋ with mounting strap

0 kPa, since no seal possibility

Process temperature (transducer tem-

perature)

-40 … +80 °C (-40 … +176 °F)

Vibration resistance

mechanical vibrations with 4 g and 5 … 100 Hz

8)

Electromechanical data - version IP 66/IP 67 and IP 66/IP 68; 0.2 bar
Cable entry/plug

9)

Ʋ Single chamber housing

– 1 x cable gland M20 x 1.5 (cable: ø 5 … 9 mm), 1 x

blind plug M20 x 1.5

or:

– 1 x closing cap M20 x 1.5; 1 x blind plug M20 x 1.5

or:

– 1 x closing cap ½ NPT, 1 x blind plug ½ NPT

or:

– 1␣x plug (depending on the version), 1␣x blind stopper

M20␣x␣1.5

Ʋ Double chamber housing

– 1 x cable entry M20 x 1.5 (cable: ø 5 … 9 mm), 1 x

blind plug M20 x 1.5; 1 x blind plug M16 x 1.5 or

optionally available with 1 x plug M12 x 1 for external

display and adjustment unit

or:

– 1 x closing cap ½ NPT, 1 x blind plug ½ NPT, 1 x blind

plug M16 x 1.5 or optionally 1 x plug M12 x 1 for exter-

nal display and adjustment unit

or:

– 1 x plug (depending on the version), 1 x blind plug

M20 x 1.5; 1 x blind plug M16 x 1.5 or optionally avail-

able with 1 x plug M12 x 1 for external display and

adjustment unit

Spring-loaded terminals for wire cross-

section

< 2.5 mm² (AWG 14)

Display and adjustment module
Voltage supply and data transmission

through the sensor

Indication

LC display in dot matrix

Adjustment elements

4 keys

Protection rating

Ʋ unassembled

IP 20

Ʋ mounted into the sensor without cover IP 40

Material

Ʋ Housing

ABS

8)

Tested according to the guidelines of German Lloyd, GL directive 2.

9)

Depending on the version M12 x 1, according to DIN 43650, Harting, 7/8" FF.

Advertising