Code

Message
type/NA
64

Message/De-
scription

Cause

Measure

Priority

702 (W702) Warning C

C> Memory
module data
faulty

l

The data were not
written correctly into
the memory module,
e.g. if the memory
module was removed
during a writing oper-
ation

l

Data memory module
contains no data

l

Repeat upload

l

Carry out reset (Code
7864) and make a
fresh adjustment

l

Copy suitable data in-
to the data memory
module (see operating
instructions manual
VEGADIF 55, chapter
"C

opy configuration

data")

55

703 (W703) Alarm B

B> Measure-
ment transmis-
sion error

l

Fault in the main elec-
tronics

l

Main electronics de-
fective

l

Separate instrument
briefly from voltage
supply

l

Exchange the main
electronics

22

704 (W704) Alarm B

B> Measure-
ment transmis-
sion error

l

Fault in the main elec-
tronics

l

Main electronics de-
fective

l

Separate instrument
briefly from voltage
supply

l

Exchange the main
electronics

12

705 (W705) Alarm B

B> Measure-
ment transmis-
sion error

l

Fault in the main elec-
tronics

l

Main electronics de-
fective

l

Separate instrument
briefly from voltage
supply

l

Exchange the main
electronics

12

10

Service instructions VEGADIF 55