Code

Message
type/NA
64

Message/De-
scription

Cause

Measure

Priority

744 (A744)

Alarm B

B> Main elec-
tronics defective

l

Electromagnetic influ-
ences are higher than
the specifications in
chapter "Technical
data
" (see product in-
formation manual VE-
GADIF 55)

l

Main electronics de-
fective

l

Restart the instrument.
Carry out a reset
(Code 62)

l

Block electromagnetic
effects or remove the
source of interference

l

Exchange the main
electronics

11

745 (A745)

Warning C

C> Sensor infor-
mation unknown

l

Sensor does not fit the
instrument (electronic
sensor type label), in-
strument continues the
measurement

l

Replace the sensor
with another suitable
one

56

746 (A746)

Warning C

C> New initiali-
sation of the
sensor

l

Electromagnetic influ-
ences are higher than
the specifications in
chapter "Technical
data
" (see product in-
formation manual VE-
GADIF 55), in general
this message is only
displayed briefly

l

Over or low pressure
exists

l

Wait several minutes

l

Restart the instrument.
Carry out a reset
(Code 7864)

l

Block electromagnetic
effects or remove the
source of interference

l

Reduce or increase
pressure

26

747 (A747)

Alarm B

B> Sensor soft-
ware and elec-
tronics not com-
patible

l

Sensor does not fit the
instrument (electronic
sensor type label)

l

Replace the sensor
with another suitable
one

16

748 (A748)

Alarm B

B> Memory error
in the signal pro-
cessor

l

Electromagnetic influ-
ences are higher than
the specifications in
chapter "Technical
data
" (see product in-
formation manual VE-
GADIF 55)

l

Main electronics de-
fective

l

Block electromagnetic
effects or remove the
source of interference

l

Exchange the main
electronics

15

However, should these measures not be successful, call the
VEGA service hotline in urgent cases under the phone no. +49
1805 858550

.

The hotline is available to you 7 days a week round-the-clock.
Since we offer this service world-wide, the support is only
available in the English language. The service is free of
charge, only the standard telephone costs will be charged.