VEGA VEGADIF 55 Service instructions User Manual

Page 13

Advertising
background image

Code

Message
type/NA
64

Message/De-
scription

Cause

Measure

Priority

719 (A719)

Alarm B

B> Y-value of the
linearization
chart outside the
editing limits

l

At least one Y-value of
the linearization chart
is below the MIN.
TANK CONTENT or
above the MAX. TANK
CONTENT

l

Carry out a fresh ad-
justment (see operat-
ing instructions man-
ual VEGADIF 55)

l

Check the parameter
adjustment and carry
out a fresh download

40

720 (A720)

Error C
factory set-
ting warn-
ing

C> Sensor sub-
normal tempera-
ture

l

The temperature
measured in the sen-
sor is lower than the
lower nominal temper-
ature of the sensor
(see operating in-
structions manual VE-
GADIF 55, parameter
description Tmin.
SENSOR)

l

Unsuitable download
carried out

l

Loose contact on the
sensor cable

l

Increase process/am-
bient temperature

l

Check the parameter
adjustment and carry
out a fresh download

l

Wait a short moment
and ensure a stable
connection or avoid
loose connection.

40

721 (A721)

Alarm B

B> ZERO POINT

level outside the
editing limits

l

LEVEL MIN. or LEVEL
MAX. was changed

l

Carry out reset (Code
2710) and make a
fresh adjustment

41

722 (A722)

Alarm B

B> ADJUST-
MENT EMPTY

or ADJUST-
MENT FULL

outside the edit-
ing limits

l

LEVEL MIN. or LEVEL
MAX. was changed

l

Carry out reset (Code
2710) and make a
fresh adjustment

42

723 (A723)

Alarm B

B> MAX. FLOW
outside its edit-
ing limits

l

FLOW TYPE was
changed

l

Carry out a fresh ad-
justment

43

725 (A725)

Alarm B

B> Sensor con-
nection error,
cycle interfered

l

Electromagnetic influ-
ences are higher than
the specifications in
chapter "Technical
data
" (see product in-
formation manual VE-
GADIF 55)

l

Sensor or main elec-
tronics defective

l

Block electromagnetic
effects or remove the
source of interference

l

Exchange sensor or
main electronics

25

Service instructions VEGADIF 55

13

Maintenance and fault rectification

31732

-

EN

-080430