VEGA VEGADIF 55 Service instructions User Manual

Page 8

Advertising
background image

Code

Message
type/NA
64

Message/De-
scription

Cause

Measure

Priority

121 (A121)

Alarm B

B> Checksum
error in the pro-
duction EE-
PROM

l

Main electronics de-
fective

l

Exchange the main
electronics

5

122 (A122)

Alarm B

B> Checksum
error in the pro-
duction EE-
PROM

l

Cable connection,
sensor main elec-
tronics interrupted

l

Electromagnetic influ-
ences are higher than
the specifications in
chapter "Technical
data
" (see product in-
formation manual VE-
GADIF 55)

l

Main electronics de-
fective

l

Sensor defective

l

Check cable connec-
tion and repair if nec-
essary

l

Block electromagnetic
effects or remove the
source of interference

l

Exchange the main
electronics

l

Exchange the sensor

13

130 (A30)

Alarm B

B> EEPROM is
defective

l

Main electronics de-
fective

l

Exchange the main
electronics

10

131 (A131)

Alarm B

B> Checksum
error in the edit-
ing limits EE-
PROM

l

Main electronics de-
fective

l

Exchange the main
electronics

9

132 (A132)

Alarm B

B> Checksum
error in the sum
counter EE-
PROM

l

Main electronics de-
fective

l

Exchange the main
electronics

7

133 (A133)

Alarm B

B> Checksum
error in the his-
tory EEPROM

l

An error occurred dur-
ing a writing operation

l

Main electronics de-
fective

l

Carry out reset (Code
7864) and make a
fresh adjustment

l

Exchange the main
electronics

8

602 (W602) Warning C

C> Linearization
curve not monot-
onically increas-
ing

l

The linearization chart
is not monotonically
increasing or falling

l

Complete or correct
linearization chart,
then accept the linea-
rization chart again

57

8

Service instructions VEGADIF 55

Maintenance and fault rectification

31732

-EN

-080430