VEGA VEGADIF 55 Service instructions User Manual

Page 5

Advertising
background image

1

3

t

I

max.

min.

3.6 mA

21...23 mA

Fig. 1: Current output in case of alarm
1

Max. alarm (110 %): adjustable via the parameter SET MAX. ALARM
between 21 … 23 mA

2

Hold measured value: The last measured value is held

3

Min. alarm (-10 %): 3.6 mA

Default setting: OUTPUT FAIL MODE = Max. Alarm (110 %),
SET MAX. ALARM = 22 mA

With the parameter ALT. CURR. OUTPUT you set the current
outut value for the error message E120 "Sensor low pressure"
and E115 "Sensor gauge pressure". The following options are
available:

l

normal: The current output takes on the value which was
set via the parameters OUTPUT FAIL. MODE and SET
MAX. ALARM.

l

NAMUR

-

Decreasing of the lower sensor limit (E120 "Sensor low
pressure
"): 3.6 mA

-

Exceeding of the upper sensor limit (E115 "Sensor
gauge pressure
"): Current output takes on the value set
via the parameter SET MAX. ALARM.

Factory setting: ALT. CURR. OUTPUT = normal

Depending on the settings for the parameters ALARM DISPL.
TIME and ACK. ALARM MODE, the following measures must
be carried out so that a messages extinguishes:

Adjustments

1

)

Measures

l

ALARM DISPL. TIME = 0 s

l

ACK. ALARM MODE = off

l

Remove cause of the message. See also supplementary instructions
manual "Service instructions VEGADIF 55"

l

ALARM DISPL. TIME > 0 s

l

ACK. ALARM MODE = off

l

Remove cause of the message. See also supplementary instructions
manual "Service instructions VEGADIF 55"

l

Wait for alarm display time

1)
1)

Menu path for ALARM DISPL. TIME and ACK. ALARM MODE: GROUP
SELECTION - OPERATING MENU - DIAGNOSTICS - MESSAGES.

Confirmation of mes-
sages

Service instructions VEGADIF 55

5

Maintenance and fault rectification