8 dismounting – VEGA VEGAMET 381 Ex User Manual

Page 30

Advertising
background image

30

8 Dismounting

VEGAMET 381 Ex • 4 … 20 mA signal conditioning instrument

30418-01-131025

Fehlercode Ursache

Beseitigung

E017

Adjustment span

too small

Carry out a fresh adjustment and increase the distance between

min. and max. adjustment

E021

Scaling span too

small

Carry out a fresh scaling, increase the distance between min. and

max. scaling.

E110

Relay switching

points too close

together

Increase the difference between the two relay switching points

7.3 How to proceed in case of

repair

You can find a repair form as well as de-

tailed information on how to proceed under

www.vega.com/downloads and "Forms and

certificates".
By doing this you help us carry out the repair

quickly and without having to call back for nee-

ded information.
If a repair is necessary, please proceed as

follows:

Print and fill out one form per instrument

Clean the instrument and pack it damage-

proof

Attach the completed form and, if need

be, also a safety data sheet outside on the

packaging

Please contact the agency serving you to

get the address for the return shipment.

You can find the agency on our home page

www.vega.com.

8 Dismounting

8.1 Dismounting steps

Take note of chapters "Mounting" and "Connec-

ting to power supply" and carry out the listed

steps in reverse order.

8.2 Disposal

The instrument consists of materials which can

be recycled by specialised recycling companies.

We use recyclable materials and have designed

the parts to be easily separable.
WEEE directive 2002/96/EG

This instrument is not subject to the WEEE

directive 2002/96/EG and the respective nati-

onal laws. Pass the instrument directly on to a

specialised recycling company and do not use

the municipal collecting points. These may be

used only for privately used products according

to the WEEE directive.
Correct disposal avoids negative effects on

humans and the environment and ensures

recycling of useful raw materials.
Materials: see chapter "Technical data"
If you have no way to dispose of the old instru-

ment properly, please contact us concerning

return and disposal.