VEGA VEGAFLEX 66 (-110…+400°C) Profibus PA User Manual
Page 33
l
Language
l
Sensitivity
l
PA scaling unit
l
PV-Out-Scale
l
Channel
l
Additional PA value
The following safety-relevant data are not read out or written:
l
HART mode
l
PIN
l
SIL
l
Sensor length/Sensor type
l
Gating out of false signals
Copy sensor data
Copy sensor data?
Basic adjustment
If the function "Reset" is carried out, the sensor resets the
values of the following menu items to the reset values (see
chart):
2)
The following values will be reset:
Function
Reset value
Max. adjustment
Distance, upper dead zone
Min. adjustment
Distance, supplied sensor length
Min. adjustment - Cable version
Distance, lower dead zone
Integration time ti
0 s
Linearisation
linear
Sensor-TAG
Sensor
Display
PA-Out 1
Channel (PV)
lin %
Additional PA value (PV)
lin %
Out-Scale-Unit
%
PV-Out-Scale
0.00 lin % = 0.0 %
100.0 lin % = 100 %
The values of the following menu items are not reset to the
reset values (see chart) with "Reset":
2)
Sensor-specific basic adjustment.
Reset
VEGAFLEX
66 • Profibus PA
33
6 Set up with the indicating and adjustment module PLICSCOM
32764
-
EN
-080716