6 Set up with the display and adjustment module

MINITRAC 31 • 4 … 20 mA/HART - four-wire

40447-EN-131119

Additional adjustments: Instrument unit, reset, date/time, copying

function
Info: Instrument name, hardware and software version, date of manu-

facture, instrument features

Check if the correct language is already set for the display. If not, you

can change the language in the menu item "Display/Language".

Start with the setup of MINITRAC 31.
In the main menu point "Setup", the individual submenu points

should be selected one after the other and provided with the correct

parameters to ensure optimum adjustment of the measurement. The

procedure is described in the following.
Stick with the normal sequence of the menu items as closely as pos-

sible.
Setup
In this menu item you can assign an unambiguous name to the sen-

sor or the measurement loop.
This parameter is described in the operating instructions manual

"Display and adjustment module".

In this menu item you can adjust the MINITRAC 31 to the integrated

isotope in the source container.
For this purpose, check which isotope is integrated in the source

container. You can find this information on the type label of the source

container.

With this selection, the sensitivity of the sensor is adapted in an opti-

mum way to the isotope.
The MINITRAC 31 requires this information for the decay compensa-

tion. An annual recalibration is hence not necessary.
Enter the requested parameters via the appropriate keys, save your

settings with [OK] and jump to the next menu item with the [ESC] and

the [->] key.

Enter here, the respective application.
This menu item enables adaptation of the sensor to the requested ap-

plication. You can choose between the following applications: "Level",

"Density", "Limit level", "X-ray alarm" or "Real value correction".

Procedure

Setup - Measurement

loop name

Setup/Isotope

Setup/Application