7 Setup with PACTware

WEIGHTRAC 31 • 4 … 20 mA/HART - four-wire

42374-EN-131119

Carry out a setup in the sequence described in the following.

Check if the correct language is already set for the software. If not,

you can change the language in the menu item "Extras/Options".
In this menu you can assign an unambiguous name to the sensor or

the measurement loop.

After starting PACTware, you can choose if you want to carry out the

extended adjustment with the DTM (Device Type Manager) or work

with the setup assistant.

In this menu item you can first of all adjust the WEIGHTRAC 31 to the

integrated isotope in the source container.
For this purpose, check which isotope is integrated in the source

container. You can find this information on the type label of the source

container.
Through this selection, the sensitivity of the sensor is adapted

perfectly to the isotope. The normal reduction of the emitter activity is

hence considered through the radioactive decay.
The WEIGHTRAC 31 requires this information of the automatic decay

compensation. This ensures an interference-free measurement over

the complete life time of the gamma emitter - an annual recalibration

is not necessary.
Setup

Start the setup assistant by clicking the respective button.
Work as far as possible according to the sequence of the setup as-

sistant.

This menu item enables adaptation of the sensor to the requested ap-

plication. You can choose between the following applications: "Mass

flow (belt/spiral)" or "Mass flow summation slave".

Presettings

Start screen

Isotope

Setup assistant

Selection of the applica-

tion (step 1)